GUNDUA-HLANGANISA: COVID-19 DIAGNOSIS USING ASSEMBLING METHOD FOR IMAGERY

GUNDUA-HLANGANISA: COVID-19 DIAGNOSIS USING ASSEMBLING METHOD FOR IMAGERY

Structured information on the device

User and evironment

Who is the intended user? Technician
Is training required in addition to the expected skill level of the intended user? No
Is any maintenance or calibration required by the user at the time of use? Yes
Where will the technology be used? rural settings, urban settings, outdoors, indoors, primary level (health post, health center), secondary level (general hospital), tertiary level (specialist hospital)

Health technology specifications

Dimensions (mm3) 450 mm x 450 mm x 2000 mm
Weight (kg) 150 kg
Does it require the use of consumables? For example, disposable batteries, disposable electrodes, etc. Yes. X-ray film
Estimated life time 8 Years
Estimated shelf life 1 Years
Can it have a telemedicine or eHealth application? Yes
Does it use any kind of software? Yes
If yes, please describe the software, whether it is open source or proprietary, its use and/or license fee, etc. (10-70 words). The GUNDUA-HLANGANISA proprietary software is tailored to the African demographic, it uses image processing technologies on chest x-ray images to identify COVID 19 infected patients. GUNDUA-HLANGANISA proprietary software is available as a desktop software or cloud hosted software. The revenue model is a annual subscription fee for each user/client.
If yes, can the software be customized for local use? Please explain (10-70 words). Include languages available. The GUNDUA-HLANGANISA software is based on a trained convolutional neural network. The ability to localise the software depends on access to quality local data. We have used a combination of public and proprietary quality local data. Furthermore, the software can be used in the diagnosis and analysis of other pulmonary infections post COVID19, for any local outbreak.
Is it portable? Portable
Type of use Long term use
Does the technology require maintenance?

Yes

Type: verification of functions- movement, X-ray production, image acquisition, interior cleaning, and electrical and mechanical tests

Frequency: once a year

Can it be done on-site / home / community?: Yes

If yes, who should provide maintenance? Biomedical engineering staff and/or service contract with the manufacturer or third-party organization
Energy requirements

Continuous power supply Voltage required: 0 V

Facility requirements

appropriate shielding; imaging workstations or x-ray viewboxes, capable of operating and being stored continuously in ambient temperature of 0 to 50 deg C and relative humidity of 15 to 90%